Solicitation Number:2025-001041Trade Agreement:Implementing Rules and RegulationsProcurement Mode:Negotiated Procurement – Two Failed Biddings (Sec. 53.1)Classification:GoodsCategory:Medical and Dental EquipmentApproved Budget for the Contract:PHP 553,000.00 Delivery Period:15 Day/sClient Agency:Contact Person:JOSEPH M. ERECRE
PhilGEPS Alternate In-charge
New Capitol Complex,
Tagbilaran City
Bohol
Philippines 6300
63-038-4110138 Ext.4212
63-38-09566438524
ppmu@bohol.gov.ph
StatusActive
Associated Components1
Bid Supplements2
Document Request List8
Date Published01/10/2025
Last Updated / Time14/05/2026 11:27 AM
Closing Date / Time19/05/2026 17:00 PM
Description ANNUAL PROCUREMENT PLAN (QUARTERLY)
City of Tagbilaran
FOR THE YEAR 2025 – Quarter : 2
MEDICAL AND DENTAL DEVICES AND EQUIPMENT
Province, City or Municipality: Provincial Government of Bohol
Total Cost Unit Qty Particulars Item No:
Date Submitted : Constingency Regular All (Consolidated) Department / Office :
553,000.00 Total
ALTERNATIVE MODE
NEGOTIATED
PROCUREMENT
TWO-FAILED
BIDDINGS
Mode of Procurement:
553,000.00 Planned Amount 2025-001041 Plan Control No.
PROVINCIAL GOVERNMENT OF BOHOL
2025-001041
General Diagnostics Devices & Equipment Lot No.
1
Unit 1
Newborn Screening Hearing Machine – testing of the
newborn baby;s ability to hear. newborn screening for
hearing is done with automated brainstem response
tests or less often, with what are called oto-acoustic
emission or conventional auditory brainstem response
tests, DOH approved
with LED monitor 47cm/19 monitor
CPU, keyboard with mouse and AVR
Vital Sign Monitor – Features:
8.4” color TFT LCD Screen (touch screen is an
optional)
Portable, lighter weight and sturdy design
MEWS(Modified Early Warning Score)
Calculations: Hemodynamics/Dose Calculation
Rechargeable Li-ion Battery (up to 15 hours
uninterruptible work)
Spot-check and continuous monitoring mode
Selectable for Adult, pediatric and neonatal patients
Wire/WIFI/3G/4G CMS, Support HL7 protocol to HIS
Barcode scanner, infrared ear thermometer support
Night mode selectable, helps to reduce light stimulation
and noise
Graphical and tabular trend review(120 hours)
48 hours full disclosure wave review for each patient
Display
8.4” color TFT LCD Screen,relosution:600×800
ECG
Lead type
3-lead:l, II, Ill
S· lead:l, II, lll,aVR, a VI., aVF, V
ECG waveform:1 channel,7 channels
Display sensitivity:
2.5mm;mv (x0.25), 5mm;mv (x0.5), 10mm;mv (x1.0),
20mm;mv {x2.0)
Wave sweep speed: 6.25mm/s, 12.5 mm/s, 25 mm/s,
50 mm/s
Bandwidth
Diagnostic mode: 0.05Hz-100Hz
Monitor mode: 0.5Hz-40Hz
Surgery mode: 1Hz-20Hz
Strong filter mode: 5Hz-20Hz
ANNUAL PROCUREMENT PLAN (QUARTERLY)
City of Tagbilaran
FOR THE YEAR 2025 – Quarter : 2
MEDICAL AND DENTAL DEVICES AND EQUIPMENT
Province, City or Municipality: Provincial Government of Bohol
Total Cost Unit Qty Particulars Item No:
Date Submitted : Constingency Regular All (Consolidated) Department / Office :
553,000.00 Total
ALTERNATIVE MODE
NEGOTIATED
PROCUREMENT
TWO-FAILED
BIDDINGS
Mode of Procurement:
553,000.00 Planned Amount 2025-001041 Plan Control No.
PROVINCIAL GOVERNMENT OF BOHOL
2025-001041
General Diagnostics Devices & Equipment Lot No.
1
Unit 1
Newborn Screening Hearing Machine – testing of the
newborn baby;s ability to hear. newborn screening for
hearing is done with automated brainstem response
tests or less often, with what are called oto-acoustic
emission or conventional auditory brainstem response
tests, DOH approved
with LED monitor 47cm/19 monitor
CPU, keyboard with mouse and AVR
1
370,000.00 Lot/ SubCategory Total :
Patient Monitoring System Devices & Equipment Lot No.
2
PGBh-US Procurement unit 1
CMRR>100dB
Notch: 50/60Hz notch filter can be set to on or off
Differential input impedance> 5M?
Electrode polarization voltage range: ±400mV
HR range:15-350 bpm
Baseline recovery time<3s after defibrillation {in monitor
and
surgery mode)
Calibration signal:1mV (peak – peak), accuracy ±3%
RESP
Measurement method : Thoracic electrical
bioimpedance
Measuring lead: Lead I, II
Wave gain: x0.25, xO.5, x 1, x2
Respiratory impedance range: 0.5 -5?
0-150 bpm
Baseline impedance: 500-4000?
Gain: 10 grades
Scan speed: 6.25mm/s, 12.5 mm/s, 25mm/s
TEMP
Measurement method: Thermistor
Measuring range: 5~5O degree celcius (41-122degree
farenheit)
Resolution: 0.1degree celcius
Measurement accuracy: ±0.1degree celcius
Recorder (optional)
Built-in, Thermal dot array
Horizontal resolution :16 dots/mm (25 mm/s paper
speed)
Vertical resolution:8 dots/mm
Paper speed:25 mm/s, 50 mm/s
Number of waveform channels:3
NIBP
Measurement method : Automatic oscillometric method
Operating mode: Manual, automatic, continuous
Measurement unit: mmHg/kPa selectable
Typical measurement time: 20-35s
Measurement type: Systolic, Diastolic, Mean
Measurement range {mmHg)
Range of Systolic pressure:
Adult 40-270
Pediatric 40-230
Neonatal 40-135
Range of Diastolic pressure:
Adult 10-215
Pediatric 10-150
Neonatal 10-100
Range of Mean pressure:
Measurement accuracy
Adult 20-235
Pediatric 20-165
Neonatal 20-110
Measurement accuracy
Maximum average error: ±5mmHg
Maximum standard deviation: 8mmHg
Resolution: 1 mmHg
lnterval:1 ,2,3,4,5, 10, 15,30,60,90, 120,
180,240,480minutes
Overpressure protection: Software and hardware,
double
safety protection
Cuff pressure range: 0-290mmHg
SpO2
Measurement range : 0-100%
Resolution: 1%
Parameter monitoring: perfusion index(PI):0-20%
optional pleth variability index(PVI)
Accuracy: ±2% or ±2bpm
Refreshing Rate: 1 s
Pulse rate
Range: 30~300 bpm
Resolution: 1bpm
Accuracy: ±2bpm(non motion)
1
PGBh-US Procurement
±5bpm (motion)
Refreshing rate:1s
Operating Environment
Input power: AC 100-250V,50/60Hz
Temperature: 5-40 degree celcius
Humidity: <85%
Patient Range: Adult, Pediatric, Neonate
183,000.00 Lot/ SubCategory Total :